Abstract
We investigated how performance trade-off depends on dual-row transmit array geometry, position of a head inside the array and array tuning condition. A clear winner was not found. For overlapped arrays, different tuning conditions resulted in larger peak-to-peak variation for CP excitation, but B1+v decrease was as a bit less for excitations that minimized IB1+v as compared with nonoverlapped arrays. For non-overlapped array the safety excitation efficiency was better if the distance between the crown of the head and the end of the array was small.
Keywords
"Geometry","Tuning","Safety","Nonhomogeneous media","Load modeling","Magnetic resonance imaging","Power measurement"