The platelet hammer: In vitro platelet activation under repetitive hypershear
Author
Jawaad Sheriff;Phat L. Tran;Marcus Hutchinson;Tracy DeCook;Marvin J. Slepian;Danny Bluestein;Jolyon Jesty
Author_Institution
Department of Biomedical Engineering, Stony Brook University, NY 11794 USA
fYear
2015
Firstpage
262
Lastpage
265
Abstract
Mechanical circulatory support (MCS) devices, such as ventricular assist devices and the total artificial heart, have emerged as a vital therapy for advanced and end-stage heart failure. However, MCS patients face life-long antiplatelet and anticoagulant therapy to minimize thrombotic complications resulting from the dynamic and supraphysiologic device-associated shear stress conditions, whose effect on platelet activation is poorly understood. We repeatedly exposed platelets to average shear stresses up to 1000 dyne/cm2 for as short as 25 ms in the “platelet hammer,” a syringe-capillary viscometer. Platelet activation state was measured using a modified prothrombinase assay and morphological changes analyzed using scanning electron microscopy. An increase in stress accumulation (SA), the product of shear stress and exposure time, led to an increase in the platelet activation state and post-high shear platelet activation rate, or sensitization. A significant increase in platelet activation state was observed beyond an SA of 1500 dyne-s/cm2, with a marked increase in pseudopod length visible beyond an SA of 1000 dyne-s/cm2. The platelet hammer may be used to study other shear-dependent pathologies and may ultimately enhance the safety and effectiveness of MCS devices and objective antithrombotic pharmacotherapy management.
Keywords
"Stress","Pathology","Shape","Heart","Valves","Blood"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7318350
Filename
7318350
Link To Document