DocumentCode
3684158
Title
Image based quantitative reader for Lateral flow immunofluorescence assay
Author
Kaushik Basak Chowdhury;Jayaraj Joseph;Mohanasankar Sivaprakasam
Author_Institution
Department of Electrical Engineering, Indian Institute of Technology Madras, India
fYear
2015
Firstpage
1223
Lastpage
1226
Abstract
Fluorescence Lateral flow immunoassays (LFIA) have wide range of applications in point-of-care testing (POCT). An integrated, motion-free, accurate, reliable reader that performs automated quantitative analysis of LFIA is essential for POCT diagnosis. We demonstrate an image based quantitative method to read the lateral flow immunofluorescence test strips. The developed reader uses line laser diode module to illuminate the LFIA test strip having fluorescent dye. Fluorescence light coming from the region of interest (ROI) of the LFIA test strip was filtered using an emission filter and imaged using a camera following which images were processed in computer. A dedicated control program was developed that automated the entire process including illumination of the test strip using laser diode, capturing the ROI of the test strip, processing and analyzing the images and displaying of results. Reproducibility of the reader has been evaluated using few reference cartridges and HbA1c (Glycated haemoglobin) test cartridges. The proposed system can be upgraded to a compact reader for widespread testing of LFIA test strips.
Keywords
"Strips","Fluorescence","Cameras","Optical filters","Diode lasers","Hardware","Image processing"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7318587
Filename
7318587
Link To Document