• DocumentCode
    3684191
  • Title

    Creating eye closure in patients with facial nerve paralysis using an implantable solenoid actuator

  • Author

    Shaheen Hasmat;Nigel H. Lovell;Timothy Eviston;Rafael Ekmejian;Gregg J. Suaning;Jonathan Clark

  • Author_Institution
    Univ. of New South Wales, Australia
  • fYear
    2015
  • Firstpage
    1357
  • Lastpage
    1360
  • Abstract
    This paper proposes the use of an implantable solenoid actuator to create a more natural eyelid closure over current lid loading therapies in patients with facial nerve paralysis (FNP). The actuator works by moving a magnet when a solenoid is activated. This is used to tension a sling applied to the upper eyelid which closes the eye. The sling design has been described elsewhere and creating eye closure using it requires a force of 627 (± 128) mN over a movement of approximately 6 mm. The actuator described here was able to successfully achieve these parameters and repeatedly perform eyelid closure in a cadaveric rabbit model. Device limitations and future improvements have also been identified and discussed.
  • Keywords
    "Solenoids","Force","Actuators","Eyelids","Capacitors","Muscles","Implants"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7318620
  • Filename
    7318620