Title :
A smart pill for drug delivery with sensing capabilities
Author :
R. Goffredo;D. Accoto;M. Santonico;G. Pennazza;E. Guglielmelli
Author_Institution :
Biomedical Robotics and Biomicrosystems Lab, Università
Abstract :
In this paper a novel system for local drug delivery is described. The actuation principle of the micropump used for drug delivery relies on the electrolysis of a water-based solution, which is separated from a drug reservoir by an elastic membrane. The electrolytically produced gases pressurize the electrolytic solution reservoir, causing the deflection of the elastic membrane. Such deflection, in turn, forces the drug out of its reservoir through a nozzle. The proposed system is integrated in a swallowable capsule, equipped with an impedance sensor useful to acquire information on the physiological conditions of the tissue. Such information can be used to control pump activation.
Keywords :
"Impedance","Impedance measurement","Drugs","Reservoirs","Electrodes","Drug delivery","Electrical resistance measurement"
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
Electronic_ISBN :
1558-4615
DOI :
10.1109/EMBC.2015.7318621