• DocumentCode
    3684623
  • Title

    Electroencephalogram measurement from the hairy part of the scalp using polymer-based dry microneedle electrodes

  • Author

    M. Arai;Y. Kudo;N. Miki

  • Author_Institution
    Mechanical Engineering Department, Keio University, Yokohama, Japan
  • fYear
    2015
  • Firstpage
    3165
  • Lastpage
    3168
  • Abstract
    This paper reports a successful electroencephalogram (EEG) measurement from the hairy part of the scalp using a polymer-based dry microneedle electrode. The electrode consists of 25 pillars, each of which has a sharp microneedle on the top. Hairs are collected into the gaps of the pillars and the microneedles can reach the scalp surface. Since the microneedles can penetrate through the stratum corneum, no conductive gel is necessary to acquire high quality EEG. We experimentally investigated the pillar diameters in EEG measurement from the occipital region with hairs. The fabricated electrodes successfully measured EEG without any skin preparation or conductive gel.
  • Keywords
    "Electrodes","Electroencephalography","Hair","Scalp","Arrays","Films","Polymers"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7319064
  • Filename
    7319064