DocumentCode :
3684657
Title :
Effect of intra-aortic balloon pump on coronary blood flow during different balloon cycles support∶ A computer study
Author :
Thin Pa Pa Aye;Zwe Lin Htet;Thamvarit Singhavilai;Phornphop Naiyanetr
Author_Institution :
Department of Biomedical Engineering, Faculty of Engineering, Mahidol University, Thailand
fYear :
2015
Firstpage :
3303
Lastpage :
3306
Abstract :
Intra-aortic balloon pump (IABP) has been used in clinical treatment as a mechanical circulatory support device for patients with heart failure. A computer model is used to study the effect on coronary blood flow (CBF) with different balloon cycles under both normal and pathological conditions. The model of cardiovascular and IABP is developed by using MATLAB SIMULINK. The effect on coronary blood flow has been studied under both normal and pathological conditions using different balloon cycles (balloon off; 1:4; 1:2; 1:1). A pathological heart is implemented by reducing the left ventricular contractility. The result of this study shows that the rate of balloon cycles is related to the level of coronary blood flow.
Keywords :
"Heart","Mathematical model","Blood flow","Computational modeling","Pathology"
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN :
1094-687X
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/EMBC.2015.7319098
Filename :
7319098
Link To Document :