DocumentCode
3684781
Title
Simple method for adaptive filtering of motion artifacts in E-textile wearable ECG sensors
Author
Tamador Alkhidir;Andrzej Sluzek;Murat Kaya Yapici
Author_Institution
Electrical and Computer Engineering Department, Khalifa University of Science, Technology and Research (KUSTAR), Abu Dhabi, 127788, UAE
fYear
2015
Firstpage
3807
Lastpage
3810
Abstract
In this paper, we have developed a simple method for adaptive out-filtering of the motion artifact from the electrocardiogram (ECG) obtained by using conductive textile electrodes. The textile electrodes were placed on the left and the right wrist to measure ECG through lead-1 configuration. The motion artifact was induced by simple hand movements. The reference signal for adaptive filtering was obtained by placing additional electrodes at one hand to capture the motion of the hand. The adaptive filtering was compared to independent component analysis (ICA) algorithm. The signal-to-noise ratio (SNR) for the adaptive filtering approach was higher than independent component analysis in most cases.
Keywords
"Electrocardiography","Adaptive filters","Electrodes","Signal to noise ratio","Textiles","Filtering"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7319223
Filename
7319223
Link To Document