• DocumentCode
    3684806
  • Title

    Calibration method for lumbosacral dimensions in wearable sensor system of lumbar alignment

  • Author

    Yoshio Tsuchiya;Takashi Kusaka;Takayuki Tanaka;Yoshikazu Matsuo;Makoto Oda;Tsukasa Sasaki;Tamotsu Kamishima;Masanori Yamanaka

  • Author_Institution
    Hokkaido Univ., Japan
  • fYear
    2015
  • Firstpage
    3909
  • Lastpage
    3912
  • Abstract
    Anteflexion of the spine is essential for many physical activities in everyday life. However, this motion places the lumbar disks under heavy load due to changes in the shape of the lumbar spine and can lead to low back pain. With the aim of reducing low back pain, here we developed a wearable sensor system that can estimate lumbosacral alignment and lumbar load by measuring the shape of the lumbar skin when the lumbosacral alignment changes. In addition, we used this system to measure the parameters of anteflexion and studied the change in dimensions of the lumbar spine from changes in posture. By determining the dimensions of the lumbosacral spine on an X-ray image, a lumbosacral dimensions calibration method based on body surface area and height was developed. By using this method, lumbosacral alignment and lumbar load could be accurately estimated using the wearable sensor system.
  • Keywords
    "Estimation","Skin","X-ray imaging","Accelerometers","Wearable sensors","Back","Spine"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7319248
  • Filename
    7319248