Title :
Medial temporal lobe high resolution magnetic resonance images for the early diagnosis of Alzheimer´s disease
Author :
Francesco Brun;Francesco Sensi;Rossella Quartulli;Luca Rei;Alban Grucka;Valentina Mancarella;Andrea Chincarini;Maja Ukmar;Agostino Accardo;Renata Longo
Author_Institution :
Dipartimento di Ingegneria e Architettura, Università
Abstract :
A challenging point in neuroimaging is the diagnosis of Alzheimer´s disease (AD) during its asymptomatic phase. Among all the biomarkers proposed in the literature, a measure of the hippocampal atrophy via Magnetic Resonance Imaging (MRI) seems to be one of the most reliable. Refined image processing techniques were already proposed to automatically extract the hippocampal boxes from images acquired with the standard full brain acquisition protocol suggested by the Alzheimer´s Disease Neuroimaging Initiative (ADNI). In order to enhance this approach, here we propose a high resolution (HR) MRI protocol focused on the medial temporal lobe (MTL) mainly conceived for 1.5T MRI device, hereafter referred as MTL-HR protocol. A preliminary characterization of its behavior when compared to the standard ADNI protocol is also presented.
Keywords :
"Magnetic resonance imaging","Protocols","Image resolution","Alzheimer´s disease","Standards","Bandwidth"
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
Electronic_ISBN :
1558-4615
DOI :
10.1109/EMBC.2015.7319339