Characterization and reduction of motion artifacts in photoplethysmographic signals from a wrist-worn device
Author
Alexandra-Maria Tăuţan;Alex Young;Eva Wentink;Fokko Wieringa
Author_Institution
Holst Centre/imec High Tech Campus 31, 5656 AE Eindhoven, The Netherlands
fYear
2015
Firstpage
6146
Lastpage
6149
Abstract
Methods for analyzing various motion artifacts in photoplethysmography (PPG) signals, recorded by a wristworn device are reported. The analysis looks both at intrinsic PPG signal properties, through standard deviation, skew and kurtosis, but also at its relationship to five possible motion reference signals, through the correlation coefficient and mutual information. The investigated references include the X, Y, Z accelerometer axes, acceleration magnitude and a different wavelength PPG channel. The latter showed higher correlation to the input PPG signal during movement. A wavelet based motion artifact reduction algorithm is described. The algorithm improves the visibility of the heart rate (HR) components in the frequency domain representation of the signals.
Keywords
"Wrist","Correlation","Sensors","Standards","Mutual information","Accelerometers"