DocumentCode :
3685345
Title :
Characterization and reduction of motion artifacts in photoplethysmographic signals from a wrist-worn device
Author :
Alexandra-Maria Tăuţan;Alex Young;Eva Wentink;Fokko Wieringa
Author_Institution :
Holst Centre/imec High Tech Campus 31, 5656 AE Eindhoven, The Netherlands
fYear :
2015
Firstpage :
6146
Lastpage :
6149
Abstract :
Methods for analyzing various motion artifacts in photoplethysmography (PPG) signals, recorded by a wristworn device are reported. The analysis looks both at intrinsic PPG signal properties, through standard deviation, skew and kurtosis, but also at its relationship to five possible motion reference signals, through the correlation coefficient and mutual information. The investigated references include the X, Y, Z accelerometer axes, acceleration magnitude and a different wavelength PPG channel. The latter showed higher correlation to the input PPG signal during movement. A wavelet based motion artifact reduction algorithm is described. The algorithm improves the visibility of the heart rate (HR) components in the frequency domain representation of the signals.
Keywords :
"Wrist","Correlation","Sensors","Standards","Mutual information","Accelerometers"
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN :
1094-687X
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/EMBC.2015.7319795
Filename :
7319795