• DocumentCode
    3685408
  • Title

    Peripheral venous blood oxygen saturation can be non-invasively estimated using photoplethysmography

  • Author

    Musabbir Khan;Christopher G. Pretty;Alexander C. Amies;Rodney B. Elliott;Fatanah M. Suhaimi;Geoffrey M. Shaw;J. Geoffrey Chase

  • Author_Institution
    Dept. of Mechanical Engineering, University of Canterbury, Private Bag 4800, Christchurch 8140, New Zealand
  • fYear
    2015
  • Firstpage
    6405
  • Lastpage
    6408
  • Abstract
    Measurement of peripheral venous oxygen saturation (SvO2) is currently performed using invasive catheters or direct blood draw. The purpose of this study was to non-invasively determine SvO2 using a variation of pulse oximetry techniques. Artificial respiration-like modulations applied to the peripheral vascular system were used to infer regional SvO2 using photoplethysmography (PPG) sensors. To achieve this modulation, an artificial pulse generating system (APG) was developed to generate controlled, superficial perturbations on the finger using a pneumatic digit cuff. These low pressure and low frequency modulations affect blood volumes in veins to a much greater extent than arteries due to significant arterial-venous compliance differences. Ten healthy human volunteers were recruited for proof-ofconcept testing. The APG was set at a modulation frequency of 0.2 Hz (12 bpm) and 45-50 mmHg compression pressure. Initial analysis showed that induced blood volume changes in the venous compartment could be detected by PPG. Estimated arterial oxygen saturation (97% [IQR=96.1%-97.4%]) matches published values (95%-99%). Estimated venous oxygen saturation (93.2% [IQR=91.-93.9%]) agrees with reported ranges (92%-95%) measured in peripheral regions. The median difference between the two saturations was 3.6%, while the difference between paired measurements in each subject was statistically significant (p=0.002). These results demonstrate the feasibility of this method for real-time, low cost, non-invasive estimation of SvO2. Further validation of this method is warranted.
  • Keywords
    "Blood","Fingers","Modulation","Band-pass filters","Finite impulse response filters","Sensors","Veins"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7319858
  • Filename
    7319858