• DocumentCode
    3685528
  • Title

    Neutron distribution and induced activity inside a Linac treatment room

  • Author

    B. Juste;R. Miró;G. Verdú;S. Díez;J.M. Campayo

  • Author_Institution
    Instituto de Seguridad Industrial, Radiofí
  • fYear
    2015
  • Firstpage
    6896
  • Lastpage
    6899
  • Abstract
    Induced radioactivity and photoneutron contamination inside a radiation therapy bunker of a medical linear accelerator (Linac) is investigated in this work. The Linac studied is an Elekta Precise electron accelerator which maximum treatment photon energy is 15 MeV. This energy exceeds the photonuclear reaction threshold (around 7 MeV for high atomic number metals). The Monte Carlo code MCNP6 has been used for quantifying the neutron contamination inside the treatment room for different gantry rotation configuration. Walls activation processes have also been simulated. The approach described in this paper is useful to prevent the overexposure of patients and medical staff.
  • Keywords
    "Neutrons","Photonics","Linear particle accelerator","Monte Carlo methods","Head","Geometry","Linear accelerators"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7319978
  • Filename
    7319978