DocumentCode
3685528
Title
Neutron distribution and induced activity inside a Linac treatment room
Author
B. Juste;R. Miró;G. Verdú;S. Díez;J.M. Campayo
Author_Institution
Instituto de Seguridad Industrial, Radiofí
fYear
2015
Firstpage
6896
Lastpage
6899
Abstract
Induced radioactivity and photoneutron contamination inside a radiation therapy bunker of a medical linear accelerator (Linac) is investigated in this work. The Linac studied is an Elekta Precise electron accelerator which maximum treatment photon energy is 15 MeV. This energy exceeds the photonuclear reaction threshold (around 7 MeV for high atomic number metals). The Monte Carlo code MCNP6 has been used for quantifying the neutron contamination inside the treatment room for different gantry rotation configuration. Walls activation processes have also been simulated. The approach described in this paper is useful to prevent the overexposure of patients and medical staff.
Keywords
"Neutrons","Photonics","Linear particle accelerator","Monte Carlo methods","Head","Geometry","Linear accelerators"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7319978
Filename
7319978