DocumentCode
3685622
Title
A comparison of classification based confidence metrics for use in the design of myoelectric control systems
Author
E. Scheme;K. Englehart
Author_Institution
Institute of Biomedical Engineering at the University of New Brunswick, Fredericton, Canada, E3B 5A3
fYear
2015
Firstpage
7278
Lastpage
7283
Abstract
In many pattern recognition applications, confidence scores are used to extract more information than discrete class membership alone, yet they have not traditionally been leveraged in myoelectric control. In this work, the confidence scores of eight common classification schemes were examined. Their role in rejecting inadvertent motions is investigated, and the tradeoffs observed in the design of rejection capable control schemes are demonstrated. It is shown that the distribution of confidences can varying greatly between classifiers, even when classification performance is similar. As a specific example, an ensemble of support vector machines in a one against one configuration (SVM1vs1) outperforms the previously reported LDAR myoelectric pattern recognition rejection scheme in terms of accuracy-rejection curves (ARC) and false acceptance/rejection (FAR) curves.
Keywords
"Accuracy","Training","Pattern recognition","Testing","Standards","Measurement","Robustness"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7320072
Filename
7320072
Link To Document