DocumentCode
3686104
Title
Electrical Bio-Impedance based non-invasive method for the Central Aortic blood pressure waveform estimation
Author
Andrei Krivoshei;Mart Min;Hasso Uuetoa;Jürgen Lamp;Paul Annus
Author_Institution
TJS Department of Electronics, Tallinn University of Technology, Estonia
fYear
2014
Firstpage
181
Lastpage
184
Abstract
Non-invasive and safe method for estimation of the Central Aortic Pressure (CAP) waveform is presented. It is based on the measurement of Electrical Bio-Impedance (EBI) on radial artery. Reconstruction of the CAP curve from the EBI data is provided by an introduced spectral domain transfer function (TF). The TF estimation algorithm and the Augmentation Index (AI) evaluation method are described. Proposed approach is less dependent on the operator and more convenient than traditional tonometry based method. It allows continuous and long term monitoring of the AI and CAP variations. Feasibility and accuracy of the technique was evaluated in clinical surroundings by comparing invasively measured CAP and AI data with non-invasive EBI based estimates. Clinical experiments were conducted in the Heart Centre of the East-Tallinn Central Hospital during scheduled coronary angiography. Age of the patients varied randomly between 43 and 80.
Keywords
"Artificial intelligence","Transfer functions","Estimation","Blood pressure","Arteries","Pressure measurement","Biomedical monitoring"
Publisher
ieee
Conference_Titel
Electronic Conference (BEC), 2014 14th Biennial Baltic
Type
conf
DOI
10.1109/BEC.2014.7320586
Filename
7320586
Link To Document