DocumentCode :
3686555
Title :
Flow and temperature control for an innovative hyperthermic intraperitoneal chemotherapy equipment
Author :
Iulia Clitan;Vlad Muresan;Mihail Abrudean;Daniel Moga;Honoriu Valean;Corneliu Lungoci;Ion Aurel Mironiuc
Author_Institution :
Department of Automation, Technical University of Cluj-Napoca, Romania
fYear :
2015
Firstpage :
162
Lastpage :
167
Abstract :
Nowadays cytoreductive surgery followed by hyperthermic intraperitoneal chemotherapy (HIPEC) represents a standard treatment for Peritoneal Carcinomatosis, a final stage of abdominal cancer. HIPEC requires the use of an equipment that delivers the cytostatic solution inside the peritoneal cavity at high temperature of 41°C-43°C. In this paper the authors present a temperature control structure for an innovative HIPEC equipment, that is under development. A discrete Internal Model Controller is designed for the cytostatic solution´s temperature control. Due to the fact that the heating process depends on the time the solution spends inside the heating element a feed-forward control structure in respect to the solution flow is also analyzed.
Keywords :
"Transfer functions","Process control","Mathematical model","Heat transfer","Resistance heating"