Title :
Phase calibration of oscilloscopes at low frequencies
Author :
Joo-Gwang Lee;Chihyun Cho
Author_Institution :
Center for Electromagnetic Wave, Korea Research Institute of Standards and Science, Daejeon, Republic of Korea
Abstract :
This paper describes how to calibrate phase response of oscilloscopes, especially at low frequencies from 10 MHz to 1 GHz. The method uses the phase response of a standard mixer calibrated by three-mixer method as the reference phase. We describe basic principles and show results obtained using a balanced mixer and a real-time oscilloscope (1 GHz BW).
Keywords :
"Oscilloscopes","Mixers","Calibration","Standards","Transfer functions","Clocks","Microwave measurement"
Conference_Titel :
Infrared, Millimeter, and Terahertz waves (IRMMW-THz), 2015 40th International Conference on
Electronic_ISBN :
2162-2035
DOI :
10.1109/IRMMW-THz.2015.7327761