• DocumentCode
    3692167
  • Title

    Multi-focus tissue harmonic images obtained with parallel transmit beamforming by means of orthogonal frequency division multiplexing

  • Author

    Libertario Demi;Gabriele Giannini;Alessandro Ramalli;Piero Tortoli;Massimo Mischi

  • Author_Institution
    Lab. of Biomed. Diagnostics, Eindhoven Univ. of Technol., Eindhoven, Netherlands
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Parallel beamforming in transmit (PBT) by means of orthogonal frequency division multiplexing (OFDM) was recently applied to increase the frame rate of ultrasound tissue harmonic imaging. OFDM-PBT improves the data acquisition rate by generating, for each transmission event, multiple beams that are allocated to a specific portion of the available transducer bandwidth. Improving the frame rate is however not the only possible application of an increased data acquisition rate. In this paper, OFDM-PBT is exploited to perform multi-focus tissue harmonic imaging for improving the penetration depth and signal to noise ratio (SNR) without affecting the frame rate. Tissue harmonic images of a tissue mimicking phantom were obtained with OFDM-PBT and standard B-mode imaging. Results present improved penetration depth and SNR for OFDM-PBT as compared to standard B-mode. We estimated an improvement of 6 dB of the average SNR, at the expenses of a reduction of the axial resolution (0.7 vs 1.1 mm).
  • Keywords
    "OFDM","Imaging","Standards","Signal to noise ratio","Harmonic analysis","Ultrasonic imaging","Image resolution"
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2015 IEEE International
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2015.0501
  • Filename
    7329155