DocumentCode
3692306
Title
Prospective discrimination of vertebral fractures by axial transmission ultrasound using optimized first arriving signal velocity measurements
Author
J. Schneider;K. Raum;M. Pumberger;T. Zippelius;E. Hoff;P. Strube;M. Putzier;J.G. Minonzio;P. Laugier
Author_Institution
Berlin-Brandenburg School for Regenerative Therapies, Charité
fYear
2015
Firstpage
1
Lastpage
4
Abstract
The objective of this study is to assess the clinical performance of a novel bi-directional axial transmission (BDAT) ultrasound prototype. The speed of sound of the first arriving signal (VFAS) was measured at the radius (single site) and at the tibia (two sites) of 51 post-menopausal females. An optimized VFAS measurement protocol was introduced to reduce the clinical measurement time (from 30 to 10 minutes) by maintaining the robustness of the former method. The optimized protocol was efficient and showed excellent discrimination performance: the highest areas under the ROC curves (AUC) were found for VFAS radius (OR: 3.54; AUC: 0.81) and VFAS tibia medial surface (MS) (OR: 4.10; AUC: 0.81). Dual energy x-ray absorptiometry (DXA) at the lumbar spine exhibited similar fracture discrimination power (OR: 1.89; AUC: 0.81) compared to the ultrasonic method.
Keywords
"Ultrasonic variables measurement","Bones","Ultrasonic imaging","Acoustics","Velocity measurement","Bidirectional control"
Publisher
ieee
Conference_Titel
Ultrasonics Symposium (IUS), 2015 IEEE International
Type
conf
DOI
10.1109/ULTSYM.2015.0513
Filename
7329296
Link To Document