DocumentCode
3703630
Title
PD of two kinds of blood pulse wave as an alternative of PAT
Author
Yibin Li;Xiaomeng Chen;Ning Deng
Author_Institution
Institute of Microelectronics, Tsinghua University, Beijing, China
fYear
2015
Firstpage
15
Lastpage
18
Abstract
Peak Delay (PD) of two different blood pulse waves measured at the same site is intensively studied in this paper. The derived expression, based on Womersley theory and simplified Navier-Stokes equations, implies that PD is closely related to the change of the Young´s modulus of the blood vessel, which is promising in noninvasively BP monitoring. Traditionally, BP is estimated with Pulse Transit Time (PTT) or Pulse Arrival Time (PAT). Experiment results show that PAT and PD are inversely related with correlation coefficient from 0.78 to 0.95 when BP is relatively stable. If BP changes remarkably, calibration should be conducted to better trace PAT with PD. Therefore PTT or PAT can be replaced by PD for continuous and noninvasive BP monitoring, with the convenience that the measuring site is concentrated on fingertip.
Keywords
"Biomedical monitoring","Blood pressure","Monitoring","Pulse measurements","Young´s modulus","Yttrium"
Publisher
ieee
Conference_Titel
Bioelectronics and Bioinformatics (ISBB), 2015 International Symposium on
Type
conf
DOI
10.1109/ISBB.2015.7344912
Filename
7344912
Link To Document