• DocumentCode
    3704470
  • Title

    Electromagnetic techniques to minimize the risk of hazardous local heating around medical implant electrodes during MRI scanning

  • Author

    Steven McCabe;Jonathan Scott;Stephen Butler

  • Author_Institution
    Department of Engineering, The University of Waikato, Hamilton, New Zealand
  • fYear
    2015
  • Firstpage
    702
  • Lastpage
    705
  • Abstract
    Magnetic Resonance Imaging (MRI) scans are contraindicated for many patients with medical implants. We establish the circumstances that cause, and the resistances required to ameliorate and to eliminate dangerous levels of MRI-induced heating that occur at the exposed, distal end of an electrical lead implanted in tissue. Simulated predictions are compared with measurements made at 128 MHz in a 3-Tesla MRI machine. A low resistance at kilohertz frequencies is sought by implant makers, in contrast with the high resistance demanded for safety. The practicality of presently-developed strategies to prevent tissue damage is brought into question. We examine the extent to which skin-depth and transmission-line properties can be manipulated to improve safety.
  • Keywords
    "Wires","Magnetic resonance imaging","Insulation","Implants","Electrodes","Resistance","Temperature measurement"
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference (EuMC), 2015 European
  • Type

    conf

  • DOI
    10.1109/EuMC.2015.7345860
  • Filename
    7345860