  • Title

    Q-factor change of cavity by sample installation in resonance perturbation method

  • Author

    Taro Miura;Ken Tahara;Jun-ichi Sugiyama;Masahiro Horibe

  • Author_Institution
    The Mueller, JAPAN
  • fYear
    2015
  • Firstpage
    738
  • Lastpage
    741
  • Abstract
    In the measuring process of the cavity resonance perturbation method to evaluate permittivity, it has historically been supposed that Q-factor of the cavity itself is not affected by the installation of a sample since the method has been proposed. Recently, several experimental results to show the change of Q-factor by the installation of a sample were found. Measurements and discussions to confirm the change have been made and a method to obtain the intrinsic loss factor of the sample separated from Q-factor change of the cavity is proposed in this paper.
  • Keywords
    "Cavity resonators","Q-factor","Loss measurement","Tuners","Current density","Resonant frequency","Dielectric loss measurement"
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference (EuMC), 2015 European
  • Type

    conf

  • DOI
    10.1109/EuMC.2015.7345869
  • Filename
    7345869