Title : 
Development of a 27.12 MHz CC-LCF intraluminal applicator for hyperthermia of the esophagus
         
        
            Author : 
J Crezee;HJ Harms;HJ Boluijt;J Sijbrands;MCCM Hulshof;HP Kok
         
        
            Author_Institution : 
Dept. Radiation Oncology, Academic Medical Center, Amsterdam, The Netherlands
         
        
        
        
        
            Abstract : 
Esophageal tumors can be treated with radiation combined with hyperthermia. This study investigated the design of a 27.12 MHz capacitively coupled intraluminal applicator for hyperthermia of the esophagus to be used in combination with external locoregional heating. It consists of four independent electrodes mounted on an inflatable balloon catheter, 10mm each, Measured SAR penetration depth is 4.5mm. Spatial steering is excellent and will prove useful in heterogeneous tumors. Combination with locoregional heating showed a strong increase in temperature rise. Next step is application in esophageal cancer.
         
        
            Keywords : 
"Applicators","Hyperthermia","Electrodes","Temperature measurement","Tumors","Heating","Temperature distribution"
         
        
        
            Conference_Titel : 
Microwave Conference (EuMC), 2015 European
         
        
        
            DOI : 
10.1109/EuMC.2015.7345877