• DocumentCode
    3706138
  • Title

    A 5 ?W/channel 9b-ENOB BioADC array for electrocortical recording

  • Author

    Elliot Greenwald;Ernest So;Mohsen Mollazadeh;Christoph Maier;Ralph Etienne-Cummings;Nitish Thakor;Gert Cauwenberghs

  • Author_Institution
    Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21205
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    We present a 4-channel integrated biopotential analog-to-digital converter (bioADC) for electrocortical recording in 180nm CMOS. The bioADC directly transduces microvolt biopotentials into a digital representation without a voltage-amplification stage. Each channel comprises a continuous-time first-order ΔΣ modulator with OTA input and current feedback, followed by a second-order comb-filter decimator. Powered from a 1.5V supply, the analog and digital circuits in each channel draw 2.1uA and 1.4uA of supply current, respectively. The bioADCs achieve an SNR of 57dB and a SFDR or 63.5dB, for 9-b ENOB. Intracranial EEG recordings from an anesthetized rat are shown and compared to simultaneous recordings from a commercial reference system to validate performance in-vivo. The micropower operation and direct digital readout make this circuit ideally suited for closed-loop neuromodulation applications.
  • Keywords
    "Latches","Capacitors","Current measurement","Frequency measurement","Electrodes","Power demand","Noise measurement"
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
  • Type

    conf

  • DOI
    10.1109/BioCAS.2015.7348309
  • Filename
    7348309