Title : 
Functional neuroimaging using UWB impulse radar: A feasibility study
         
        
            Author : 
Timo Lauteslager;Nicoletta Nicolaou;Tor Sverre Lande;Timothy Constandinou
         
        
            Author_Institution : 
Department of Electrical and Electronic Engineering, Imperial College London, UK
         
        
        
        
        
            Abstract : 
Microwave imaging is a promising new modality for studying brain function. In the current paper we assess the feasibility of using a single chip implementation of an ultra-wideband impulse radar for developing a portable and low-cost functional neuroimaging device. A numerical model is used to predict the level of attenuation that will occur when detecting a volume of blood in the cerebral cortex. A phantom liquid is made, to study the radar´s performance at different attenuation levels. Although the radar is currently capable of detecting a point reflector in a phantom liquid with submillimeter accuracy and high temporal resolution, object detection at the desired level of attenuation remains a challenge.
         
        
            Keywords : 
"Attenuation","Phantoms","Blood","Liquids","Neuroimaging","Ultra wideband radar"
         
        
        
            Conference_Titel : 
Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
         
        
        
            DOI : 
10.1109/BioCAS.2015.7348387