• DocumentCode
    3706234
  • Title

    A low-voltage design of controller-based ADPLL for implantable biomedical devices

  • Author

    Jungnam Bae;Saichandrateja Radhapuram;Ikkyun Jo;Takao Kihara;Toshimasa Matsuoka

  • Author_Institution
    Graduate School of Engineering, Osaka University, 2-1 Yamada-oka, Suita-shi, Osaka 565-0871, Japan
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A low-voltage controller-based all-digital phase-locked loop (ADPLL) utilized in the medical implant communication service (MICS) frequency band is designed. The controller-based loop topology is used to control the phase and frequency for reliable handling of the ADPLL output signal. The digitally-controlled oscillator with the delta-sigma modulator is employed to achieve high frequency resolution. The phase error is reduced by the phase selector with 64-phase signal from the phase interpolator. Fabricated in a 130-nm CMOS process, the ADPLL has an active area of 0.64 mm2. It consumes 840 μW from a 0.7-V supply voltage and has a settling time of 80 μs. The measured phase noise is -114.6 dBc/Hz at 200 kHz offset frequency.
  • Keywords
    "Phase frequency detector","Microwave integrated circuits","Frequency measurement","Phase noise","Tuning","Frequency conversion"
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
  • Type

    conf

  • DOI
    10.1109/BioCAS.2015.7348405
  • Filename
    7348405