DocumentCode :
3706279
Title :
Column-based micro-beamformer for improved 2D beamforming using a matrix array transducer
Author :
Hyun gil Kang;Sua Bae;Pilsu Kim;Jiwon Park;Gunho Lee;Woojin Jung;Minsuk Park;Kangsik Kim;Wooyoul Lee;Tai-Kyong Song
Author_Institution :
Department of Electronic Engineering, Sogang University, Seoul, Republic of Korea
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
In a 3-D medical ultrasound imaging system, a matrix array probe with 2-D positioning of the elements allows high resolution of ultrasound images due to its capability of two-dimensional dynamic focusing. However, the hundreds (up to thousands) of elements in the matrix array make the fabrication of the transducers and cables challenging. In this paper, to achieve high quality of 3-D ultrasound images with low hardware complexity, we introduce a column-based micro-beamformer (CMB) in which a column in the matrix array is considered as a sub-array and then elevational and lateral beamformings are sequentially conducted in the analog and digital stages, respectively. For the performance evaluation of the proposed beamformer, the beam pattern simulations are conducted and point-spread-functions are obtained. In addition, root-mean-square-errors (RMSEs) in round-trip time delays were measured over the depths. Compared to a conventional micro-beamformer, the CMB produced more tightly focused beam patterns and showed almost equivalent performance to that of fully sampled array. In the near field, the mean RMSEs of the proposed and conventional beamformers were 274 ns and 18.1 ns (improved 93.4% of delay accuracy), respectively.
Keywords :
"Array signal processing","Arrays","Ultrasonic imaging","Imaging","Delays","Probes","Transducers"
Publisher :
ieee
Conference_Titel :
Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
Type :
conf
DOI :
10.1109/BioCAS.2015.7348450
Filename :
7348450
Link To Document :