DocumentCode :
3731541
Title :
Thermal-Converter Validation of Differential Sampling Measurement Based on AC-Programmable Josephson Voltage Standard System
Author :
Y. Amagai;M. Maruyama;H. Yamamori;C. Urano;S.-F. Chen;H. Fujiki;N.-H. Kaneko
Author_Institution :
Nat. Inst. of Adv. Ind. Sci. &
fYear :
2015
fDate :
7/1/2015 12:00:00 AM
Firstpage :
1
Lastpage :
3
Abstract :
This paper describes a differential sampling measurement system using AC-programmable Josephson voltage standard (AC-PJVS) system up to 10 V in the frequency range down to 1 Hz. Such system combines the versatility and the accuracy and stability of the Josephson voltage standard. The validity of this method is confirmed with the AC-DC difference measurements of a calibrated thermal converter, and with an electro-thermal simulation at the root-mean square amplitudes of 3 V. The uncertainty of our differential sampling measurement system is estimated to be 1.3 μV/V for waveform frequencies of 62.5 Hz and root-mean square amplitude of 10 V.
Keywords :
"Voltage measurement","Frequency measurement","Uncertainty","Standards","Semiconductor device measurement","Voltmeters","Measurement uncertainty"
Publisher :
ieee
Conference_Titel :
Superconductive Electronics Conference (ISEC), 2015 15th International
Type :
conf
DOI :
10.1109/ISEC.2015.7383486
Filename :
7383486
Link To Document :
بازگشت