• DocumentCode
    3735701
  • Title

    Assembly of transducer array using anisotropic conductive film for medical imaging applications

  • Author

    Hoang-Vu Nguyen;Trym Eggen;Knut E. Aasmundtveit

  • Author_Institution
    HBV - Buskerud and Vestfold University College, Borre, Norway
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    Interconnection technology based on anisotropic conductive film (ACF) has been selected to assemble a transducer array on a flexible substrate in ultrasound imaging applications. The process of bonding and subsequently dicing a transducer on a flexible substrate was evaluated. The results show good integrity of ACF interconnects in test samples undergoing the assembly process. The interconnects, composed of a transducer element and a flex pad, retain high mechanical shear strength and low electrical resistance even after the tough dicing process. Furthermore, a high process yield promising for implementation in industry has been obtained. This work has demonstrated the feasibility of using ACF for assembling ultrasound transducers on system substrates.
  • Keywords
    "Substrates","Flexible printed circuits","Ultrasonic imaging","Transducers","Resistance","Assembly","Bonding"
  • Publisher
    ieee
  • Conference_Titel
    Microelectronics Packaging Conference (EMPC), 2015 European
  • Type

    conf

  • Filename
    7390724