DocumentCode
3741638
Title
Microfilaria filtering microfluidic chip ? Preliminary study
Author
Saktip Uthongsap;Surasak Kasetsirikul;Werayut Srituravanich;Alongkorn Pimpin;Sariya Asawakarn;Prapruddee Piyaviriyakul;Wutthinan Jeamsaksiri;Witsaroot Sripumkhai
Author_Institution
Department of Mechanical Engineering, Faculty of Engineering, Chulalongkorn University, Thailand
fYear
2015
Firstpage
1
Lastpage
4
Abstract
This paper provides a preliminary experiment of microfluidic filtering chip, designed to be used in an area that lacks of laboratory equipment. Without blood preparation involved sophisticated machines, the blood sample was used without rinsing or hemorrhaging before feeding into the chip. This chip contained with a wedge shape zone and detectable zone, which was in a cylindrical shape. The tests were performed with different sizes of inlet and length of the chip. From experimental results, the 5-μm micro capillaries allowed blood to flow through and was be able to filter a larger substance inside. However, appropriate blood dilution-ratio and filtering efficiency are needed to be clarified in the further experiment.
Keywords
"Grippers","Filtering","Microfluidics","Shape","Heart","Red blood cells"
Publisher
ieee
Conference_Titel
Biomedical Engineering International Conference (BMEiCON), 2015 8th
Type
conf
DOI
10.1109/BMEiCON.2015.7399517
Filename
7399517