DocumentCode
3741660
Title
Development of a triangular micro well for single cell trapping ? Computational study
Author
Tewan Tongmanee;Alongkorn Pimpin;Werayut Srituravanich;Wutthinan Jeamsaksiri;Witsaroot Sripumkhai;Dettachai Ketpun;Achariya Sailasuta;Prapruddee Piyaviriyakul
Author_Institution
Department of Mechanical Engineering, Faculty of Engineering Chulalongkorn University, Thailand
fYear
2015
Firstpage
1
Lastpage
4
Abstract
The aim of this study is to find appropriate parameters of a trapping device, i.e. equilateral triangular microwell array, using a computational simulation. This simulation was divided into nine cases of microwell with various depths and widths to help analyze the effects of them on velocity and vorticity distribution induced inside the microwell. The microwells with equilateral length of 40, 60 and 80μm, and each length consisted of three depths, i.e. 15, 30 and 45 μm, were studied. The results suggested that the equilateral triangular 40-μm microwell with the depth of 30 μm should be able to effectively trap a 10-20 μm particle.
Keywords
"Charge carrier processes","Arrays","Fluids","Computational modeling","Boundary conditions","Analytical models","Force"
Publisher
ieee
Conference_Titel
Biomedical Engineering International Conference (BMEiCON), 2015 8th
Type
conf
DOI
10.1109/BMEiCON.2015.7399540
Filename
7399540
Link To Document