DocumentCode
3749085
Title
Estimation of the maximal heart rate to improve online tonic-clonic seizure detection using ECG
Author
Thomas De Cooman;Anouk Van de Vel;Berten Ceulemans;Lieven Lagae;Wim Van Paesschen;Bart Vanrumste;Sabine Van Huffel
Author_Institution
Department of Electrical Engineering (ESAT), STADIUS, KU Leuven, Belgium
fYear
2015
Firstpage
977
Lastpage
980
Abstract
Previous studies have shown that during several types of seizures, the heart rate increases strongly towards a maximal patient-specific epileptic heart rate HRep. This ictal peak heart rate is one of the most important features for classifying epileptic heart rate increases. We therefore try to estimate HRep, which is done by using least squares support vector machines. The found estimation had a mean square error of 18bpm, which is an improvement compared to age-based estimators. Adding this information to an online seizure detector led to an increased performance (F1-score: 14.65% to 18.72%) with a decreased detection delay (23.8s to 11.9s).
Keywords
"Lead","Heart rate","Art","Europe","Electrocardiography","Brain modeling","Electroencephalography"
Publisher
ieee
Conference_Titel
Computing in Cardiology Conference (CinC), 2015
ISSN
2325-8861
Print_ISBN
978-1-5090-0685-4
Electronic_ISBN
2325-887X
Type
conf
DOI
10.1109/CIC.2015.7411076
Filename
7411076
Link To Document