• DocumentCode
    3749137
  • Title

    Reducing false arrhythmia alarms in the ICU using novel signal quality indices assessment method

  • Author

    Runnan He;Henggui Zhang;Kuanquan Wang;Yongfeng Yuan;Qince Li;Jiabin Pan;Zhiqiang Sheng;Na Zhao

  • Author_Institution
    School of Computer Science and Technology, Harbin Institute of Technology, China
  • fYear
    2015
  • Firstpage
    1189
  • Lastpage
    1192
  • Abstract
    The physiological signals such as the electrocardiogram (ECG) and arterial blood pressure (ABP) in the ICU are often severely corrupted by noise, artifact and missing data, producing large errors in the estimation of the characteristics of the signals values, leading to false alarms in ICU. In order to solve this problem, we started with the signal quality assessment of vital signals in intensive care patients using a derived signal quality index (SQI) to reveal the degree of signal quality. And then we use the SQI-weighted residual error of Kalman filters (KF) to complete the date fusion for evaluating the heart rate (HR). Finally, the algorithm of arrhythmia false alarm reduction in ICU monitors was developed based upon the method of combining SQIs and HR estimations derived from ECG waveform and ABP waveform recorded from ICU patients. Results show that the overall True Positive Rate (TPR), True Negative Rate (TNR) and overall score for the Event-1 are respectively 65%, 82%, and 53.19, for the Event-2, the TPR, TNR and overall score are 65%, 87%, and 54.64.
  • Keywords
    "Electrocardiography","Lead","Estimation","Physiology","Filtering","Heart rate","Patents"
  • Publisher
    ieee
  • Conference_Titel
    Computing in Cardiology Conference (CinC), 2015
  • ISSN
    2325-8861
  • Print_ISBN
    978-1-5090-0685-4
  • Electronic_ISBN
    2325-887X
  • Type

    conf

  • DOI
    10.1109/CIC.2015.7411129
  • Filename
    7411129