• DocumentCode
    3753004
  • Title

    Comparison of three anatomic sheep models for magnetic resonance safety

  • Author

    Hamza Erg?der;Samat Turdumamatov;Muhammed Erg?l;Tahir Malas

  • Author_Institution
    Electrical and Electronics Engineering, Turgut ?zal University, Ankara, Turkey
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    We compare three anatomical sheep models to assess possible heating effects of magnetic resonance imaging. We generate i) homogenous, ii) three-layered, and iii) seven-tissue models and compute specific absorption rate (SAR) and temperature increase on those models. We conclude that the homogenous model overestimates both SAR and heating whereas the three-tissue model underestimates those (although slightly) compared to more accurate seven-tissue model. However, simulations with both homogenous and three-layered models can be performed much faster and using fewer computational resources compared to more accurate models, hence, can be viable alternatives to simulations with a more accurate anatomic model.
  • Keywords
    "Magnetic resonance imaging","Computational modeling","Coils","Biological system modeling","Heating","Radio frequency","Bones"
  • Publisher
    ieee
  • Conference_Titel
    Electronics Computer and Computation (ICECCO), 2015 Twelve International Conference on
  • Type

    conf

  • DOI
    10.1109/ICECCO.2015.7416880
  • Filename
    7416880