DocumentCode :
3758298
Title :
X-ray sensitivity of poly(triarylamine) (PTAA)
Author :
Naimul Hassan;Shaikh Asif Mahmood
Author_Institution :
Department of Electrical and Electronic Engineering, Bangladesh University of Engineering and Technology, Dhaka-1000, Bangladesh
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
X-ray sensitivity of semiconducting polymer poly(triarylamine) (PTAA) based direct conversion X-ray detectors is determined for mammography application considering drift of charge carriers in the presence of deep traps. Using COMSOL Multiphysics 5.0, the continuity equations for charge carriers are solved considering proper initial and boundary conditions. For mammography, the calculated X-ray sensitivity of PTAA is 0.2690 μC/cm2R at an applied field of 10 V/μm. The X-ray sensitivity of amorphous selenium (a-Se) is also determined for mammography and compared with that of PTAA. The sensitivity of PTAA is found to be similar to that of a-Se at 10 V/μm.
Keywords :
"Photoconducting materials","Sensitivity","Electrodes","Charge carrier processes","Electric fields","X-ray imaging","Current density"
Publisher :
ieee
Conference_Titel :
Telecommunications and Photonics (ICTP), 2015 IEEE International Conference on
Type :
conf
DOI :
10.1109/ICTP.2015.7427956
Filename :
7427956