    A generalized scatter reconstruction algorithm for limited energy resolution PET detectors

  • Author

    Hongyan Sun;Stephen Pistorius

  • Author_Institution
    University of Manitoba and CancerCare Manitoba, Winnipeg, R3E 0V9 Canada
  • fYear
    2014
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    A novel PET reconstruction algorithm, which can extract the radioactive distribution using both true and scattered coincidences has been adapted to handle the limited energy resolution of detectors encountered in practice. The results show that for hot regions, incorporating scattered photons into the reconstruction, improves the image quality more than simply rejecting them as noise. For cold regions the same is true, provided the energy resolution is better than 5%.