• DocumentCode
    3759589
  • Title

    Simulation study of a pinhole SPECT system with a triple head gamma camera

  • Author

    Yuta Ichimura;Koichi Ogawa

  • Author_Institution
    Graduate School of Engineering, Hosei University, Tokyo 184-8584, Japan
  • fYear
    2014
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    The purpose of our research is to develop a new pinhole SPECT system using a triple head gamma camera system. The proposed data acquisition geometry used multi-pinhole collimators attached to each gamma camera. Acquired data were reconstructed with the OS-EM method. We evaluated the feasibility of the system with simulations. In the simulations, we evaluated three types of the pinhole collimators with the numbers of pinholes; 5, 8 and 11. And we used a brain phantom and myocardial phantom. Attenuation correction was conducted with the x-ray CT image of the corresponding slice. The results of the simulations confirmed the feasibility of our proposed multi-pinhole SPECT system with more than eight pinholes for each detector.
  • Keywords
    "Image reconstruction","Collimators","Single photon emission computed tomography","Phantoms","Myocardium","Data acquisition","Detectors"
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2014.7430822
  • Filename
    7430822