• DocumentCode
    3759612
  • Title

    Calibration and data standardization of a prototype bench-top preclinical CT

  • Author

    Richard Taschereau;Nam T. Vu;Arion F. Chatziioannou

  • Author_Institution
    Crump Institute for Molecular Imaging, UCLA School of Medicine, Department of Molecular and Medical Pharmacology, Los Angeles, CA 90095 USA
  • fYear
    2014
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    This paper reports on the geometric calibration procedure of a bench-top, high-speed, and continuous-rotation preclinical computed tomography (CT) scanner that has been developed at our institution. A full 13-parameter model of the image formation process is described. Once calibrated, the model is used to modify projection data sets - the standardization operation - so that they appear to come from an ideal system. This technique has the advantage that standardized scans can be fed to any reconstruction algorithm since no awareness of the actual system geometry is required from the reconstruction program.
  • Keywords
    "Detectors","Calibration","Computed tomography","Image reconstruction","Phantoms","Standardization","Geometry"
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2014.7430845
  • Filename
    7430845