Title :
Comparing analog and digital pulse-shape-discrimination system performance for Organic-Liquid Scintillators
Author :
Charles S. Sosa;Marek Flaska;Sara A. Pozzi
Author_Institution :
Department of Nuclear Engineering at the University of Michigan, Ann Arbor, 48104 USA
Abstract :
This work compares the PSD performance of two systems: one digital (NI PXI-5124) and one analog (Mesytec MPD4). Both systems were optimized for the best possible PSD performance, when using the same data acquisition system: PXI-5124, having 12 bit resolution, 10 V pulse-height dynamic range, and a 200 MHz sampling rate. Measurement conditions were kept identical between systems. A 5×5" Organic-Liquid Scintillator (EJ-309), coupled to a PMT (ETL-9821), = was used to measure Cf-252. FOM values helped to quantify PSD performance for both systems. The results indicate that the analog PSD system marginally outperformed the digital PSD system.
Keywords :
"Photonics","Neutrons","Data acquisition","Detectors","Pulse measurements","Cleaning","Histograms"
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
DOI :
10.1109/NSSMIC.2014.7431097