• DocumentCode
    3764379
  • Title

    MW blood sample characterization using co-axial transmission line

  • Author

    Evan Hilderbrand;Joseph Korfhagen;George J. Shaw;Altan M. Ferendeci

  • Author_Institution
    Electrical Engineering and Computing Systems, University of Cincinnati, Cincinnati, Ohio 45221
  • fYear
    2015
  • fDate
    6/1/2015 12:00:00 AM
  • Firstpage
    242
  • Lastpage
    244
  • Abstract
    Microwave Tomography (MWT) is being developed as an accurate detection of hemorrhagic strokes using microwaves. A transmission line technique is used to measure the relative permittivity of blood as a function of frequency and is presented to provide insight on an appropriate operating frequency for MWT. A response calibration method is used on a coaxial sample holder to accurately measure the scattering parameters of and characterization of liquid samples. This paper demonstrates the ability to accurately measure liquid samples by replacing the coaxial line dielectric with liquid. Experimental results from water and 99% isopropyl alcohol are compared against known values to provide verification of the coaxial airline to accurately measure blood samples in the microwave region.
  • Keywords
    "Blood","Transmission line measurements","Permittivity measurement","Frequency measurement","Permittivity","Calibration","Microwave theory and techniques"
  • Publisher
    ieee
  • Conference_Titel
    Aerospace and Electronics Conference (NAECON), 2015 National
  • Electronic_ISBN
    2379-2027
  • Type

    conf

  • DOI
    10.1109/NAECON.2015.7443075
  • Filename