DocumentCode
3771428
Title
Residual losses in a guard-ring micrometer-electrode holder for solid-disk dielectric specimens
Author
Arnold H. Scott;William P. Harris
Author_Institution
National Bureau of Standards, Washington, D. C.
fYear
1960
Firstpage
11
Lastpage
17
Abstract
Residual losses occurring in the specimen holder and in the bridge standard capacitor must be taken into account in order to make accurate bridge measurements of the dielectric losses of materials that have very low losses. Also, knowledge of the exact spacing of the electrodes is required for accurate determination of the dielectric constant of materials. To better meet these requirements, a new specimen holder has been designed and constructed. By a technique using ball reference gages, the electrodes of the holder can be adjusted to be parallel to about one micron, and the zero correction of the micrometer can be determined to ±1 micron. Using a modification of a technique used by Astin1 and others, the residual loss angles of the holder and its connecting leads and of the bridge standard capacitor were determined to ±1 or 2 microradians.
Keywords
"Loss measurement","Electrodes","Capacitors","Standards","Capacitance","Dielectric loss measurement","Bridge circuits"
Publisher
ieee
Conference_Titel
Electrical Insulation, Annual Report 1960 Conference on
Print_ISBN
978-1-5090-3109-2
Type
conf
DOI
10.1109/CEI.1960.7461663
Filename
7461663
Link To Document