• DocumentCode
    3777954
  • Title

    The realization of spectral total radiance flux scale and related calibration service at NIM

  • Author

    Zhao Weiqiang; Liu Hui; Liu Jian

  • Author_Institution
    National Institute of Metrology, Beijing, China 100029
  • Volume
    1
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    285
  • Lastpage
    289
  • Abstract
    National Institute of Metrology China (NIM) has studied and developed the facilities for spectral total radiant flux (STRF) measurement and provides related calibration services. The array-based goniospectroradiometers and an integrating sphere system have been setup. The former is used to realize the scale of the STRF, and the latter is used to implement everyday calibration service due to its fast speed. The STRF standard lamps are also developed. For the goniospectroradiometer, the measurement result has an uncertainty of 1.8% ~ 3.6% (k=2) within the spectral range of 380 nm to 780 nm. For the integrating sphere system, the measurement result has an uncertainty of 2.0% ~ 3.8% (k=2) within the spectral range of 380 nm to 780 nm.
  • Keywords
    "Standards","Light emitting diodes","Spectroradiometers","Calibration","Measurement uncertainty","Uncertainty","Light sources"
  • Publisher
    ieee
  • Conference_Titel
    Electronic Measurement & Instruments (ICEMI), 2015 12th IEEE International Conference on
  • Type

    conf

  • DOI
    10.1109/ICEMI.2015.7494174
  • Filename
    7494174