• DocumentCode
    3777998
  • Title

    A design for four-terminal-pair resistor with calculable frequency dependence

  • Author

    Tong Yazhen; Huang Xiaoding

  • Author_Institution
    Beijing Orient Institute for Measurement and Test, 100190, China
  • Volume
    1
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    402
  • Lastpage
    406
  • Abstract
    Since the AC resistance measuring meter and the bridge adopt measurement principle of the four-terminal-pair, but the resistor with calculable frequency dependence is usually the four-terminal structure. So it is necessary to design a four-terminal-pair resistor with calculable frequency dependence. This paper introduces the 100Ω coaxial resistor with calculable frequency dependence based on four-terminal-pair. It introduces the structure and parameters of the resistance sample, builds the coaxial resistance´s mathematical model and calculates the related parameters. Calculated AC-DC difference is only a few parts in 10-11 at a frequency of 1kHz, the time constant of the resistance is 3.6×10-9s. By comparing with the measuring result, both have a good consistency.
  • Keywords
    "Resistance","Wires","Immune system","Frequency dependence","Resistors","Electrical resistance measurement","Standards"
  • Publisher
    ieee
  • Conference_Titel
    Electronic Measurement & Instruments (ICEMI), 2015 12th IEEE International Conference on
  • Type

    conf

  • DOI
    10.1109/ICEMI.2015.7494219
  • Filename
    7494219