• DocumentCode
    3778103
  • Title

    Remove the time base drift of sampling oscilloscope

  • Author

    Zhu Jiangmiao; Yang Jie; Miao Jingyuan

  • Author_Institution
    College of Electronic Information &
  • Volume
    2
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    772
  • Lastpage
    776
  • Abstract
    The time base drift is a key factor that influences the sampling oscilloscope to measure the high speed pulse signal precisely, so time base drift must be removed as far as possible. Cross-correlation algorithm is an alternative method to remove time base drift effectively. First, this paper does computer simulation, then use the algorithm to do research on the original data that is collected when establishing national pulse parameters primary standard. Both the results of simulation and the research show that the cross-correlation algorithm can remove the time drift effectively.
  • Keywords
    "Time measurement","Measurement uncertainty","Standards","Biographies","Oscilloscopes","Calibration"
  • Publisher
    ieee
  • Conference_Titel
    Electronic Measurement & Instruments (ICEMI), 2015 12th IEEE International Conference on
  • Type

    conf

  • DOI
    10.1109/ICEMI.2015.7494327
  • Filename
    7494327