DocumentCode :
3778153
Title :
Study on pre-travel anisotropic characteristic of touch trigger probe and its calibration method
Author :
Bohan Zhao; Gao Feng; Li Yan; Fan Lijun
Author_Institution :
Key Laboratory of NC Machine Tools and Integrated Manufacturing Equipment of Xi´an University of Technology, Ministry of Education, 710048, China
Volume :
2
fYear :
2015
fDate :
7/1/2015 12:00:00 AM
Firstpage :
1024
Lastpage :
1028
Abstract :
Systematic error resulted from the pre-travel of the touch trigger probe is non-ignorable consititutent of inspection errors for On-Machine Measurennet (OMM). Based on analyzing the working procedure of probe, the mechanics condition of triggering mechanism under working situation is performed. Utilizing the results, a probe pre-travel mathematical model is established, which can describe the pre-travel and its anisotropic characteristic quantitatively. Then a numerical simulation of the mathematical model along the radial direction of the probe is implemented. Using a high resolution displacement sensor and high speed data acquisition instrument together with a specially designed detection circuitry, a novel pre-travel calibration method and device is set up. By means of the presented device, the pre-travel characteristics of the probe along the radial direction over 360° have been obtained. Finally a contrastive analysis between theoretical model and experimental results is carried out.
Keywords :
"Probes","Force","Calibration","Switches","Sensors","Data acquisition"
Publisher :
ieee
Conference_Titel :
Electronic Measurement & Instruments (ICEMI), 2015 12th IEEE International Conference on
Type :
conf
DOI :
10.1109/ICEMI.2015.7494377
Filename :
7494377
Link To Document :
بازگشت