Testing of infrared spectroradiometer using blackbody
Author
Wang Hongchao; Liu Hongyuan; Ying Chengping; Wu Bin; Huo Mingming; Jiang Bin
Author_Institution
The 41st Institute of China Electronics Technology Group Corporation, Qingdao 266555, China
Volume
3
fYear
2015
fDate
7/1/2015 12:00:00 AM
Firstpage
1289
Lastpage
1292
Abstract
In this paper, some important parameters for a Circular Variable Filter (CVF) based spertroradiometer are tested. A standard blackbody radition source, which is calibrated and traceable to National Institute of Standards and Technology (NIST), is adopted to accomplish the test. Interesting properties are accuracy, nonlinearity, stability and Noise Equivalent Temperature (NET). The testing results show that measurement accuracy and stability can be achieved 0.47% and 0.84%. For InSb and MCT detectors, the nonlinearity is 0.95% and 1.67% respecticely, the NET is 6.6mdeg and 23.2mdeg respecticely. This test verifies performance indicators of this spertroradiometer and provides a reliable guarantee for subsequent use, this work will also helpful for the calibration of various spectrometers.
Keywords
"Temperature measurement","Mirrors","Atmospheric measurements","Detectors","Frequency measurement","Spectroradiometers","Integrated optics"