DocumentCode :
3779247
Title :
Effect of pump characteristic (H-Q) curves on the efficiency of Total Artificial Heart
Author :
Md. Rashidul Islam;Meah Imtiaz Zulkarnain
Author_Institution :
Department of Electrical & Electronic Engineering, Kitami Institute of Technology, Hokkaido 090-0015, Japan
fYear :
2015
Firstpage :
55
Lastpage :
58
Abstract :
Nowadays, heart failure of human being is a very common occurrence all over the world. Total Artificial Heart (TAH) is an outstanding achievement of biomedical engineering where the patient´s natural heart is removed and replaced with two mechanical pumps that provide perfusion to the systemic and pulmonary circulations. This paper focuses on the effect of pump characteristic (H-Q) curves on the efficiency of TAH. To show the effect of pump parameters, a simplified model of human cardiovascular system (CVS) is developed with an electrical circuit. A numerical simulation is conducted in order to find out the different physiological pressures and pump flows in response to change of parameters related to pump performance curves. Finally, the effect of H-Q curves on the efficiency of TAH is analyzed.
Keywords :
"Heart","Physiology","Mathematical model","Integrated circuit modeling","Numerical models","Electrical engineering"
Publisher :
ieee
Conference_Titel :
Advances in Electrical Engineering (ICAEE), 2015 International Conference on
Type :
conf
DOI :
10.1109/ICAEE.2015.7506795
Filename :
7506795