Title :
Extracting the heart rate variability from an electrocardiogram sampled at a very low frequency
Author :
A.K. Barros;J. Wisbeck;N. Ohnishi
Author_Institution :
RIKEN, Inst. of Phys. & Chem. Res., Saitama, Japan
Abstract :
Our study focuses on a new method of estimating the heart rate variability (HRV) which does not need the use of R-wave detection from the electrocardiogram (ECG). Contrary to this detection method which requires a sampling frequency higher than 100 Hz, the one proposed here could calculate the HRV from an ECG signal sampled around 5 Hz with a mean error of 0.03 beats/sup -1/. The method also worked efficiently to extract the HRV from an ECG measured from a subject in an exercise condition. The overall error was smaller as we low-pass filtered the heart rate variability.
Keywords :
"Heart rate variability","Electrocardiography","Sampling methods","Frequency estimation","Signal analysis","Frequency measurement","Biomedical engineering","Heart rate detection","Low pass filters","Fluctuations"
Conference_Titel :
Computers in Cardiology, 1999
Print_ISBN :
0-7803-5614-4
DOI :
10.1109/CIC.1999.825975