• DocumentCode
    3784821
  • Title

    High sensitivity spectrometer at 3 /spl mu/m based on difference frequency generation for N/sub 2/O detection

  • Author

    G. Pesce;G. Rusciano;A. Sasso

  • Author_Institution
    Dipt. Sci. Fisiche, Univ. di Napoli Federico, Italy
  • Volume
    3
  • Issue
    2
  • fYear
    2003
  • Firstpage
    206
  • Lastpage
    211
  • Abstract
    We report the description of a laser spectrometer based on difference-frequency generation. As a pump laser, we used a master-slave diode laser configuration emitting at 785 nm, while for a signal laser, we used a diode-pumped Nd : YAG laser. The wavelength of the idler radiation is continuously tunable in the spectral region between 2.88 to 3.02 /spl mu/m. In this range lie the /spl nu//sub 1/ fundamental vibrational symmetric stretching mode of water vapor and the relatively strong combination band /spl nu//sub 1/+/spl nu//sub 3/ of nitrous oxide that are both investigated in this work. We show that both molecules can be simultaneously detected by using a simple absorption scheme and by tuning the spectrometer frequency within few gigahertz. A systematic study of pressure broadening for a N/sub 2/O line has been performed in presence of the main air constituents, i.e., nitrogen and oxygen. Finally, the estimation of the minimum detectable N/sub 2/O concentration at 500 torr of nitrogen resulted in 2.6(2) ppmV with 0.9-m absorption path-length.
  • Keywords
    "Spectroscopy","Frequency","Pump lasers","Laser modes","Absorption","Nitrogen","Laser excitation","Master-slave","Diode lasers","Neodymium"
  • Journal_Title
    IEEE Sensors Journal
  • Publisher
    ieee
  • ISSN
    1530-437X
  • Type

    jour

  • DOI
    10.1109/JSEN.2002.806208
  • Filename
    1202945