DocumentCode :
378658
Title :
Experimental and numerical study of the insert-substitution method: application to the measurement of the nonlinear parameter β of solids
Author :
Matar, O.B. ; Vila, M. ; Meulen, Francois Vander
Author_Institution :
GIP Ultrasons/LUSSI, Blois
Volume :
1
fYear :
2001
fDate :
2001
Firstpage :
709
Abstract :
The object of this work is the optimization of the insert-substitution experimental method to determine the nonlinear parameter β of solid plates. The method is based on the comparison of the measured amplitudes of the second order field pressure with and without the sample being immersed in water between the emitter and the receiver. An analytical description of the second order field based on the description of the primary beams as a superposition of Gaussian beams is presented. The model takes into account the sound attenuation and dispersion, and the receiver´s size. The choice of the ultrasonic emitter and receiver (i.e. needle point hydrophone, focused, or non focused transducer) is investigated. Linear and nonlinear ultrasonic waves generated by a focused and a non-focused transducer through a plate are measured. The experimental data is found to be in good agreement with the model predictions
Keywords :
hydrophones; ultrasonic materials testing; ultrasonic propagation; ultrasonic transducers; Gaussian beams superposition; insert-substitution method; linear ultrasonic waves; needle point hydrophone; nonlinear parameter; nonlinear ultrasonic waves; second order field pressure; solid plates; Acoustic transducers; Attenuation; Needles; Optimization methods; Predictive models; Pressure measurement; Solids; Sonar equipment; Ultrasonic transducers; Ultrasonic variables measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2001 IEEE
Conference_Location :
Atlanta, GA
Print_ISBN :
0-7803-7177-1
Type :
conf
DOI :
10.1109/ULTSYM.2001.991824
Filename :
991824
Link To Document :
بازگشت